1.1 This Recommendation specifies the metrological
and technical requirements for general-purpose, liquidin-
glass thermometers(1) intended for laboratory use as
reference or working temperature standards. The
thermometers are graduated in degrees Celsius and
divided into six accuracy classes (see 5.2.2) that apply
to thermometers operating within the temperature
range of – 200 °C to + 500 °C.
1.2 This Recommendation covers solid-stem and
enclosed-scale thermometers. In addition, such thermometers
may be complete immersion, total immersion,
or partial immersion thermometers. The performance
requirements for thermometers containing
mercury and organic liquids as thermometric liquids
are specifically covered in this Recommendation.
Requirements for thermometers containing other
liquids such as gallium and mercury-thallium could be
developed by applying the same principles.
1.3 A test procedure is given to determine if the
thermometers meet each specified requirement under
specified laboratory conditions. Such procedure
includes visual inspection and tests on permanency of
pigment, bulb stability, and determination of errors.
1.4 This Recommendation does not apply to
maximum registering thermometers, such as those
used to measure body temperature.
|
|
|
|