1.1 This Recommendation may be applied for
alanine EPR dosimetry systems used to control and
supervise any application of ionizing radiation for
industrial processing of materials and products. It does
not cover, nor does it exclude, the use of other
equivalent means of measurement or determination of
absorbed dose for such applications.
1.2 This Recommendation is intended to apply
specifically to manufacturers of alanine dosimeters, to
manufacturers of EPR spectrometers used for
dosimetry, and to national bodies responsible for the
metrological control of dosimetry systems. It also
contains information of interest to users of alanine
EPR dosimetry systems.
1.3 The metrological and technical performance
requirements of alanine EPR dosimetry systems are
covered in this Recommendation; however, the selection
and routine use of these and other dosimeters for
radiation processing applications are discussed in
other standards [1, 2]. Performance better than that
prescribed in this Recommendation may be achieved.
1.4 This Recommendation applies to dosimeters
irradiated by either photons or electrons within the
energy range of 0.1 to 28 MeV [1]. Tests of dosimeters
according to this Recommendation are specified to be
carried out at a reference temperature and humidity
within a specified absorbed dose range and absorbed
dose rate range.
|
|
|
|