Technical Committee No |
TC04 |
|
Status |
Active |
|
Category |
Metrology Sector |
|
Sub Category |
|
|
$Resources:StandardType; |
Kuwaiti Standard |
GSO …./….(1) comprises a series of documents dealing with methods for the
calibration of vibration and shock pick-ups.
This standard lays down detailed specifications for the instrumentation and
procedure to be used for primary calibration of accelerometers using Earth's
gravitation.
This standard applies to rectilinear accelerometers with zero-frequency response,
mainly of the straingauge or piezoresistive type, to servopick-ups, and to primary
standard and working pick-ups.
It is applicable for ± local Earth's gravitation at 0 Hz.
The limit of uncertainty applicable is ± 0.01 m/s2.
|
|
|
|