IS0 5347 comprises a series of documents dealing with methods for the calibration of vibration and shock
pick-ups.
This part of IS0 5347 lays down detailed specifications for the instrumentation and procedures to be used for
accelerometer resonance testing. It applies to rectilinear accelerometers of the piezoresistive, piezoelectric and
variable capacitance types in the frequency range 50 Hz to 200 kHz.
The procedures are in general contrary to those described in IS0 5347-14 which is limited to undamped
accelerometers. The frequency response of a piezoelectric accelerometer depends on the value of the (lowest)
resonance frequency of the instrument when mounted on the structure to be tested. It does not appear possible to
specify a test which will determine this frequency for all installations of a given accelerometer. This procedure
gives the accelerometer resonance frequency under a set of standard reproducible conditions, with the
understanding that the resonance frequency in actual use will in all probability be appreciably different (generally
lower, by a factor depending on the mass and compliance of the test structure and the method of attachment). The
procedure is not suitable for evaluating the mounted resonance frequency for a field of application; a suitable
method is given in IS0 5348.
|
|
|
|