1.1 This test method establishes procedures for measuring
the initial particle filtration efficiency of materials used in
medical facemasks using monodispersed aerosols.
1.1.1 This test method utilizes light scattering particle
counting in the size range of 0.1 to 5.0 µm and airflow test
velocities of 0.5 to 25 cm/s.
1.2 The test procedure measures filtration efficiency by
comparing the particle count in the feed stream (upstream) to
that in the filtrate (downstream).
1.3 The values stated in SI units or in other units shall be
regarded separately as standard. The values stated in each
system must be used independently of the other, without
combining values in any way.
1.4 The following precautionary caveat pertains only to the
test methods portion, Section 10, of this specification. This
standard does not purport to address all of the safety concerns,
if any, associated with its use. It is the responsibility of the user
of this standard to establish appropriate safety and health
practices and determine the applicability of regulatory limitations prior to use.
1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the
Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical
Barriers to Trade (TBT) Committee.
|
|
|
|