1.1 This test method covers general test procedures and
evaluation criteria for the corrosion resistance of surgical
instruments intended for reuse in surgery and fabricated from
stainless steel such as, but not limited to, those listed in
Specification F899.
1.2 Austenitic (Class 3), martensitic (Class 4), precipitation
hardenable (Class 5), and ferritic (Class 6) materials shall use
the boil test.
1.3 Ferritic (Class 6) materials with a minimum 16 %
chromium content, austenitic (Class 3), and precipitation
hardenable (Class 5) materials shall use the boil test and the
copper sulfate test.
1.4 The copper sulfate test is used to detect the presence of
metallic iron and iron oxide on the surface of materials.
1.5 The copper sulfate test is not recommended for martensitic materials. (See Note X1.1.)
1.6 The boil test is applicable to martensitic, austenitic,
ferritic, and precipitation hardenable materials to detect free
iron or any other anodic surface contaminants on stainless
steel.
1.7 Values in either inch-pound or SI are to b
|
|
|
|